联系电话:
131-6293-1356

Hausser Metallized Hemacytometer Reichert Bright-Line

品牌:hausser scientific

货号:1490 1492 1483 1475 1461

产品介绍

The Bright-Line Hemacytometer is molded from a single piece of thermal and shock-resistant glass. An H-shaped moat forms two counting areas, or plateaus. A "V" slash at the loading side of each plateau facilitates charing and reduces the possibility of overflow into the moat. Each plateau features enhanced Neubauer rulings. The ruled surface is 0.1mm below the cover glass, limiting the volume of blood or fluid over a square mm at 0.1cu.mm and over each of 400 squares (within the central square mm) to 0.00025 cu. mm. Contact of the flat, polished cover glass surfaces with cover glass supports produces an exact volume of fluid over the counting area. The difference in surface tension characteristics between the metallic surface on the chamber and the polished cover glass assures smooth capillarity for precise loading and more even cell distribution.


Catalog # Description
1490 Counting chamber only
1492 Counting chamber with 2 cover glass
1483 Complete set
1475 Phase Counting Chamber w/2 cover glass (used with Phase microscope)
1461 Replacement Cover Glass

Directions for Use

Bright-Line / Dark-Line Counting Chambers Catalog Numbers: 3100, 3110, 3200, 3500, 1490, 1492, 1475, & 1483

Usage: Cell Counts

Cell Depth: 0.100mm +/- 2% (1/10mm)
Volume: 0.1 Microliter
Ruling Pattern: Improved Neubauer, 1/400 Square mm

Rulings cover 9 square millimeters. Boundary lines of the Neubauer ruling are the center lines of the groups of three. (These are indicated in the illustration below.) The central square millimeter is ruled into 25 groups of 16 small squares, each group separated by triple lines, the middle one of which is the boundary. The ruled surface is 0.10mm below the cover glass, so that the volume over each of the 16 small squares is .00025 cubic mm.

The number of cells per cubic millimeter = 
         Number of cells counted per square millimeter X dilution (if used) X 10


The number of cells per milliliter = 
        Number of cells counted per square millimeter X dilution (if used) X 10,000


One (1) Milliliter = 1000 cubic millimeters (cu mm)


One (1) Microliter (ul) = One (1) cubic millimeter (cu mm)


To clean the counting chamber: After completing the count, remove the cover glass and clean the counting chamber with water or a mild cleaning solution (10% solution of bleach). Dry the counting chamber with a soft cloth or wipe, or rinse with acetone.

合作客户

联系电话:13162931356

固定电话:0571-89925175

公司地址:杭州市西湖区金蓬街366号2号楼506,512室

Copyright 2019 杭州晓柚生物科技有限公司. 浙ICP备17006713号-1